Screen messages – Physio-Control LIFEPAK 15 User Manual
Page 256
![background image](/manuals/313537/256/background.png)
B-2
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Screen Messages
CHARGING TO XXX J
Appears when CHARGE is pressed on the front panel or standard
paddles.
CHECK FOR PULSE
AED prompt after each standard 3-shock sequence or NO SHOCK
ADVISED
message.
CHECK PATIENT!
A potentially shockable rhythm has been detected when the VF/
VT alarm is on.
CHECK PATIENT. IF NO PULSE,
PUSH ANALYZE
A potentially shockable rhythm has been detected when using
Advisory Monitoring.
CHECK PRINTER
The printer door is open, there is no paper in the printer, or
another printer malfunction exists.
CO2 AUTOZERO
EtCO
2
monitor is automatically performing a zero-point
calibration.
CO2 FILTERLINE BLOCKAGE
EtCO
2
FilterLine tubing is kinked or clogged; the message
appears after 30 seconds of unsuccessful purging.
CO2 FILTERLINE OFF
EtCO
2
FilterLine tubing is disconnected or is not securely
connected to the device.
CO2 FILTERLINE PURGING
EtCO
2
FilterLine tubing is kinked or clogged with liquid.
CO2 INITIALIZING
EtCO
2
monitor is performing a self-check.
CONNECT CABLE
Therapy cable is not connected when you press CHARGE, PACER,
or ANALYZE.
CONNECT CHEST LEADS
A 12-lead ECG analysis was requested and precordial leads are
not connected to the patient.
CONNECT ECG LEADS
ECG electrodes or leads are disconnected.
CONNECT ELECTRODES
Therapy electrodes are disconnected.
CONNECTED TO
The device is connected via Bluetooth technology to another
Bluetooth-enabled device. The name of the connected device
follows this message.
CONNECTING TO
The device is establishing communication with another
Bluetooth-enabled device. The name of the target device follows
this message.
CPR: ADULT-AIRWAY X:Y
An option for CPR metronome. The patient is an adult for whom
an advanced airway has been established. The specified C:V
ratio will be used.
CPR: ADULT-NO AIRWAY X:Y
An option for CPR metronome. The patient is an adult for whom
an advanced airway has not been established. The specified C:V
ratio will be used.
Table B-1 Summary of Screen Messages (Continued)
MESSAGE