Basic orientation – Physio-Control LIFEPAK 15 User Manual
Page 29
![background image](/manuals/313537/29/background.png)
BASIC ORIENTATION
3
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
3-11
Figure 3-8 Connectors for Temperature Monitoring Configuration
CONNECTOR
ACTION
CO2
Connect: Open CO
2
port door, insert FilterLine connector, and
turn clockwise until connector is firmly seated.
Disconnect: Rotate FilterLine connector counterclockwise and
pull connector out.
SpO2/
SpCO/
SpMet
Connect: Align cable connector with SpO
2
port and push in
until connector clicks into place.
Disconnect: Press the gray buttons on each side of the cable
connector simultaneously and pull connector out.
NIBP
Connect: Insert NIBP tubing connector into the NIBP port.
Disconnect: Press the latch on the left side of the port and pull
tubing connector out.
ECG
Connect: Align the green ECG connector with the ECG port;
position the white line on the cable facing left. Insert the
cable connector into the port until the connector is firmly
seated.
Disconnect: Pull the ECG connector straight out.
TEMP
Connect: Align the temperature adapter cable connector with
the TEMP port. Insert the cable connector into the port until
the connector is firmly seated.
Disconnect: Grip the connector and pull straight out.
CO2
SpO2
NIBP
ECG
TEMP