Electromagnetic emissions, Essential performance, Electromagnetic compatibility guidance – Physio-Control LIFEPAK 15 User Manual
Page 271: Appendix

APPENDIX
D
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
D-1
Electromagnetic Compatibility Guidance
Electromagnetic Emissions
Essential Performance
The LIFEPAK 15 monitor/defibrillator maintains safe and effective performance of the defibrillation
therapy and patient monitoring functions when operated in the electromagnetic environment
specified in Table D-2 through Table D-4.
Table D-1 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The LIFEPAK 15 monitor/defibrillator is intended for use in the electromagnetic environment specified
below. The customer or the user of the LIFEPAK 15 monitor/defibrillator should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The LIFEPAK 15 monitor/defibrillator uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The LIFEPAK 15 monitor/defibrillator is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable