4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-9
Figure 4-3 Limb Lead Electrode Placement
4. Prepare the patient’s skin for electrode application:
• Shave excessive hair at electrode site.
• For oily skin, clean skin with alcohol pad.
• Gently scrape skin to remove surface layer of dead cells and improve conduction of
electrical signals.
• Avoid locating electrodes over tendons and major muscle masses.
• Clean and dry the skin.
5. Apply ECG electrodes:
• Confirm that the package is sealed and the Use By date is not passed.
• Attach an electrode to each of the lead wires.
• Grasp electrode tab and peel electrode from carrier.
• Inspect electrode gel and make sure gel is intact (discard electrode if gel is not intact).
• Hold electrode taut with both hands. Apply the electrode flat to the skin. Smooth tape
outwardly. Avoid pressing the center of the electrode.
• Secure the trunk cable clasp to the patient’s clothing.
Note:
Electrode quality is critical for obtaining an undistorted ECG signal. Always check the
date code on electrode packages for expiration date before using on a patient. Do not use
electrodes that have expired. Disposable electrodes are intended for a single use.
6. Select the desired ECG lead on the monitor screen.
7. If necessary, adjust ECG size for accurate heart rate counting.
8. Press
to obtain an ECG printout.
AHA Labels
IEC Labels
RA
Right Arm
R
Right
LA
Left Arm
L
Left
*RL
Right Leg
N
Negative
LL
Left Leg
F
Foot
*Note: Not used for 3-lead cable.
LA/L
RA/R
LL/F
RL/N