beautypg.com

Code summary report, Patient records and reports – Physio-Control LIFEPAK 15 User Manual

Page 168

background image

7-4

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Patient Records and Reports

CODE SUMMARY Report

The LIFEPAK 15 monitor/defibrillator automatically stores a CODE SUMMARY report as part of the

Patient Record for each patient. The CODE SUMMARY report can be set up to always print in a

particular format. The available formats are shown in Table 7-2. For CODE SUMMARY setup

information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device.

To generate a CODE SUMMARY report, press

CODE SUMMARY

. If you interrupt printing of a CODE

SUMMARY report, the entire CODE SUMMARY report is reprinted when printing is resumed. “Code

Summary Complete” prints immediately following the last waveform event.

Continuous
Report

2

Provides real-time waveform data, acquired when the
device is powered on and electrodes are connected or
other waveform data is displayed in channels 2 or 3. Only
for post-event review with CODE-STAT or DT EXPRESS
software.

X

1

Transmission of a 12-lead ECG report automatically includes transmission of the Vital Signs Summary.

2

To obtain CPR analytics using CODE-STAT software, the patient’s ECG must be monitored using

PADDLES

lead in Channel 1.

Note:

All reports that are transmitted to the LIFENET System include the following information:

• Battery status

• Power adapter status

• Device usage information

• Manufacturing configuration settings

• 3:00 A.M. self-test results

Table 7-2 CODE SUMMARY Formats

FORMAT

ATTRIBUTES

Long format

• Preamble
• Event/vital sign log
• Event waveforms
• 12-lead ECG reports
• Trend Summary

Medium format

• Preamble
• Event/vital sign log
• Event waveforms
• Trend Summary

Short format

• Preamble
• Event/vital sign log
• Trend Summary

Table 7-1 Report Types (Continued)

REPORT TYPE

DESCRIPTION

PRINT

FROM

MONITOR

TRANSMIT