General Troubleshooting Tips
Fuel gauge on battery does
not illuminate
Extremely depleted battery
• Charge battery in Station-Mobile or
REDI-CHARGE battery charger.
Faulty battery
• Replace battery.
Device turns off
unexpectedly
High power draw
• Press ON immediately to turn device
back on.
Low battery power
• Replace battery immediately.
• Press ON to turn device back on.
RF equipment too close to
defibrillator
• Separate RF equipment from
defibrillator. See "Separation
Distances" on page D-4.
• Press ON to turn device back on.
Cellular equipment too close
to installed battery
• Move cellular equipment away from
installed battery.
• Press ON to turn device back on.
• If device does not turn on, replace
battery.
LIFENET Gateway (modem)
too close to installed battery
• Store modem in side pouch of
defibrillator. Do not store modem in
back pouch.
• Press ON to turn device back on.
• If device does not turn on, replace
battery.
Device won’t turn off
ON
not pressed long enough
to turn off device
• Press and hold ON for at least two
seconds.
Monitor/defibrillator
operates, but screen is
blank
Operating temperature is too
low or too high
• Operate defibrillator within specified
ambient temperature range.
Screen not operating
properly
• Print ECG strip to assess rhythm and
other active vital signs.
• Press ANALYZE and use AED mode, if
necessary.
• Contact qualified service technician.
Monitor/defibrillator
operates, but screen not
readable
Screen in direct sunlight
• Change screen from color to black
and white.
• Reposition or shield device.
• Print ECG strip to assess rhythm and
other active vital signs.
• Press ANALYZE and use AED mode, if
necessary.
Table 10-2 General Troubleshooting Tips (Continued)
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION