Specifications and performance characteristics – Physio-Control LIFEPAK 15 User Manual
Page 240
A-6
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
CO
2
Accuracy
CO
2
partial pressure at
sea level:
Accuracy:
(0–80 bpm)* 0 to 38 mmHg
(0 to 5.1 kPa)
±2 mmHg
(0.27 kPa)
39 to 99 mmHg
(5.2 to 13.2 kPa)
±5% of reading + 0.08% for every 1 mmHg
(0.13 kPa) above 38 mmHg (5.1 kPa)
(>80 bpm)*
0 to 18 mmHg
(0 to 2.4 kPa)
±2 mmHg
(0.27 kPa)
19 to 99 mmHg
(2.55 to 13.3 kPa)
±4 mmHg (0.54 kPa) or ±12% of reading,
whichever is higher
*For RR > 60 bpm, to achieve specified CO
2
accuracy, the Microstream
®
FilterLine
®
H Set for infant must be used.
Respiration Rate
Accuracy
0 to 70 bpm: ±1 bpm
71 to 99 bpm: ±2 bpm
Respiration Rate Range
0 to 99 breaths/minute
Rise Time
190 msec
Response Time
3.3 seconds (includes delay time and rise time)
Initialization Time
30 seconds (typical), 10-180 seconds
Ambient Pressure
Automatically compensated internally
Optional Display
Waveform
CO
2
pressure
Scale factors
Autoscale, 0–20 mmHg (0–4 Vol%),0–50 mmHg (0–7 Vol%),
0–100 mmHg (0–14 Vol%)
INVASIVE PRESSURE
Transducer Type
Strain-gauge resistive bridge
Transducer Sensitivity
5
V/V/mmHg
Excitation Voltage
5 Vdc
Connector
Electro Shield CXS 3102A 14S-6S
Bandwidth
Digital filtered, DC to 30 Hz (< -3db)
Zero Drift
1 mmHg/hr without transducer drift
Zero Adjustment
±150 mmHg including transducer offset
Numeric Accuracy
±1 mmHg or 2% of reading, whichever is greater, plus transducer error
Pressure Range
-30 to 300 mmHg, in six user selectable ranges
Invasive Pressure Display Display: IP waveform and numerics
Units: mmHg
Labels: P1 or P2, ART, PA, CVP, ICP, LAP (user selectable)
Certification
IEC 60601-2-34 and IEC 60601-2-49 compliant using Sentra Medical
Devices transducer kit part number IBP P 40000
(www.sentramedical.com)
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)