A-6

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Specifications and Performance Characteristics

CO

2

Accuracy

CO

2

partial pressure at

sea level:

Accuracy:

(0–80 bpm)* 0 to 38 mmHg

(0 to 5.1 kPa)

±2 mmHg
(0.27 kPa)

39 to 99 mmHg
(5.2 to 13.2 kPa)

±5% of reading + 0.08% for every 1 mmHg
(0.13 kPa) above 38 mmHg (5.1 kPa)

(>80 bpm)*

0 to 18 mmHg
(0 to 2.4 kPa)

±2 mmHg
(0.27 kPa)

19 to 99 mmHg
(2.55 to 13.3 kPa)

±4 mmHg (0.54 kPa) or ±12% of reading,
whichever is higher

*For RR > 60 bpm, to achieve specified CO

2

accuracy, the Microstream

®

FilterLine

®

H Set for infant must be used.

Respiration Rate
Accuracy

0 to 70 bpm: ±1 bpm
71 to 99 bpm: ±2 bpm

Respiration Rate Range

0 to 99 breaths/minute

Rise Time

190 msec

Response Time

3.3 seconds (includes delay time and rise time)

Initialization Time

30 seconds (typical), 10-180 seconds

Ambient Pressure

Automatically compensated internally

Optional Display
Waveform

CO

2

pressure

Scale factors

Autoscale, 0–20 mmHg (0–4 Vol%),0–50 mmHg (0–7 Vol%),
0–100 mmHg (0–14 Vol%)

INVASIVE PRESSURE

Transducer Type

Strain-gauge resistive bridge

Transducer Sensitivity

5

V/V/mmHg

Excitation Voltage

5 Vdc

Connector

Electro Shield CXS 3102A 14S-6S

Bandwidth

Digital filtered, DC to 30 Hz (< -3db)

Zero Drift

1 mmHg/hr without transducer drift

Zero Adjustment

±150 mmHg including transducer offset

Numeric Accuracy

±1 mmHg or 2% of reading, whichever is greater, plus transducer error

Pressure Range

-30 to 300 mmHg, in six user selectable ranges

Invasive Pressure Display Display: IP waveform and numerics

Units: mmHg
Labels: P1 or P2, ART, PA, CVP, ICP, LAP (user selectable)

Certification

IEC 60601-2-34 and IEC 60601-2-49 compliant using Sentra Medical
Devices transducer kit part number IBP P 40000
(www.sentramedical.com)

Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)

CHARACTERISTIC

DESCRIPTION