Appendix – Physio-Control LIFEPAK 15 User Manual
Page 259
APPENDIX
B
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
B-5
PUSH ANALYZE
Press ANALYZE to begin ECG analysis.
PUSH AND HOLD SHOCK BUTTON!
The defibrillator is in Sync mode, fully charged, and ready to
provide therapy.
PUSH AND HOLD PADDLE
BUTTONS TO SHOCK!
The defibrillator is in Sync mode, fully charged, and ready to
provide therapy with hard paddles connected.
PUSH SHOCK BUTTON!
The defibrillator is fully charged and ready to provide therapy.
PX NOT ZEROED
Transducer is connected or reconnected and is not zeroed.
PX TRANSDUCER NOT DETECTED
IP transducer is disconnected from the monitor/defibrillator.
PX ZERO FAILED
The device was unable to zero the pressure transducer.
PX ZEROED
Transducer successfully zeroed.
PX ZEROING
Monitor is establishing a zero reference.
RA LEADS OFF
ECG electrode “RA” is disconnected.
REPLACE BATTERY X
Power loss for the battery in well X is imminent.
SEARCHING FOR DEVICES
Device is attempting to identify available Bluetooth devices.
SELECT BIPHASIC ENERGY / XXX J
ENERGY SELECT
was pressed on front panel or on standard
paddles.
SELF TEST FAILED
Device detected internal error; remove device from service.
SELF TEST FAILED.
TRANSMITTING...
Device detected internal error and is transmitting test results.
Remove device from service after transmission is complete.
SELF TEST IN PROGRESS
Device is performing a self test after turning on.
SELF TEST PASSED
Device passed internal test and is available for use.
SELF TEST PASSED.
TRANSMITTING...
Device passed internal test and is transmitting test results.
SHOCK ADVISED!
The defibrillator has analyzed the patient ECG rhythm and
detected a shockable ECG rhythm.
SPCO: POOR QUALITY SIGNAL
Device is not receiving sufficient input from sensor.
SPMET: POOR QUALITY SIGNAL
Device is not receiving sufficient input from sensor.
SPO2: CHECK SENSOR
The SpO
2
sensor connection to device or application to patient
needs checked.
SPO2: LOW PERFUSION
Patient has a weak pulse.
SPO2: NO SENSOR DETECTED
A sensor is disconnected from the monitor.
SPO2: POOR QUALITY SIGNAL
Device is not receiving sufficient input from sensor.
Table B-1 Summary of Screen Messages (Continued)
MESSAGE
DESCRIPTION