Cleaning the Device
Clean the LIFEPAK 15 monitor/defibrillator, therapy and ECG cables, and batteries with a damp
sponge or cloth. Use only the cleaning agents listed below:
• Quaternary ammonium compounds
• Isopropyl alcohol
• Peracetic (peroxide) acid solutions
Note:
Carefully clean the connector ports. Do not allow cleaning fluids to penetrate the exterior
surfaces of the device.
Clean the carrying case accessory as follows and as described on its instruction tag:
• Hand wash using mild soap or detergent and water. A scrub brush may be useful for heavily
soiled spots. Cleaners such as Formula 409
®
are helpful for grease, oil, and other tough stains.
For information about cleaning the reusable monitoring sensors and cables, see the individual
monitoring section.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Do not clean any part of this device or its accessories with bleach, bleach dilution, or
phenolic compounds. Do not use abrasive or flammable cleaning agents. Do not
attempt to sterilize this device or any accessories unless otherwise specified in
accessory operating instructions.