beautypg.com

Electromagnetic immunity, Electromagnetic compatibility guidance – Physio-Control LIFEPAK 15 User Manual

Page 272

background image

D-2

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Electromagnetic Compatibility Guidance

Electromagnetic Immunity

Table D-2 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The LIFEPAK 15 monitor/defibrillator is intended for use in the electromagnetic environment specified
below. The customer or the user of the LIFEPAK 15 monitor/defibrillator should assure that it is used in
such an environment.

Immunity Test

IEC 60601

Test Level

Compliance Level

Electromagnetic Environment -

Guidance

Electrostatic
discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.

Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power
supply lines

±1 kV for input/
output lines

Not applicable

±1 kV for input/
output lines

Mains power quality should be that of a
typical commercial or hospital
environment.

Surge

IEC 61000-4-5

±1 kV line(s) to
line(s)

±2 kV line(s) to earth

Not applicable

Mains power quality should be that of a
typical commercial or hospital
environment.

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 sec

Not applicable

Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
LIFEPAK 15 monitor/defibrillator
requires continued operation during
power mains interruptions, it is
recommended that the LIFEPAK 15
monitor/defibrillator be powered from
an uninterruptible power supply or a
battery.

Power frequency
(50/60 Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.

Note: U

T

is the AC Mains voltage prior to application of the test level.