Motion detected, Electrodes or therapy cable off, Therapy – Physio-Control LIFEPAK 15 User Manual
Page 127
THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-13
Subsequent analysis for
SHOCK ADVISED
and
NO SHOCK ADVISED
sequences are the same as
described above. The energy level for Shock 2, 3, and greater depends on the
ENERGY PROTOCOL
setup and the analysis decision. When a
NO SHOCK ADVISED
decision follows a shock, the energy
level does not increase for the next shock. When a
SHOCK ADVISED
decision follows a shock, the
energy level increases for the next shock.
Motion Detected
Electrodes or Therapy Cable Off
AED Mode
If the AED detects motion during the ECG
analysis, the MOTION DETECTED, STOP MOTION
prompts occur, followed by a warning tone.
Analysis is inhibited until the motion stops
or for up to 10 seconds. After the motion
ceases or 10 seconds have elapsed, analysis
continues to completion even if motion is
still present. For possible causes of motion
detection and suggested solutions, see
Table 5-1 on page 5-19.
Motion detected!
Stop motion!
AED Mode
Connect electrodes
If therapy electrodes are not connected, the
CONNECT ELECTRODES
prompts occur until the
patient is connected.
AED Mode
Connect cable
If the therapy cable is not connected to the
defibrillator, the CONNECT CABLE message
appears until the cable is connected.