Specifications and performance characteristics – Physio-Control LIFEPAK 15 User Manual
Page 238

A-4
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
COMMUNICATIONS
The device is capable of transferring data records by wired or wireless connection. This device
complies with Part 15 of the FCC rules, and its operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
Serial Port
RS232 communication
+12V available
Bluetooth
®
technology
Bluetooth technology provides short-range wireless communication with
other Bluetooth-enabled devices.
MONITOR
ECG
ECG is monitored via several cable arrangements. A 3-wire cable is used
for 3-lead ECG monitoring. A 5-wire cable is used for 7-lead ECG
monitoring. A 10-wire cable is used for 12-lead ECG acquisition. When
the chest electrodes are removed, the 10-wire cable functions as a 4-wire
cable. Standard paddles or QUIK-COMBO pacing/defibrillation/ECG
electrodes are used for paddles lead monitoring.
Frequency Response
Monitor—0.5 to 40 Hz or 1 to 30 Hz
Paddles—2.5 to 30 Hz
12-lead ECG diagnostic—0.05 to 150 Hz
Lead Selection
Leads I, II, III (3-wire ECG cable)
Leads I, II, III, AVR, AVL, and AVF acquired simultaneously (4-wire ECG
cable)
Leads I, II, III, AVR, AVL, AVF, and C lead acquired simultaneously
(5-wire ECG cable)
Leads I, II, III, AVR, AVL, AVF, V1, V2, V3, V4, V5, and V6 acquired
simultaneously (10-wire ECG cable)
ECG Size
4, 3, 2.5, 2, 1.5, 1, 0.5, 0.25 cm/mV (fixed at 1 cm/mV for 12-lead)
Heart Rate Display
20–300 bpm digital display
Accuracy:
±
4% or
±
3 bpm, whichever is greater
QRS Detection Range
Duration: 40 to 120 msec
Amplitude: 0.5 to 5.0 mV
Common Mode Rejection
(CMRR)
ECG Leads: 90 dB at 50/60 Hz
SpO
2
/SpCO/SpMet
Sensors
Masimo
®
sensors including Rainbow™ sensors
Nellcor
®
sensors when used with the Masimo Red™ MNC adapter
SpO
2
Displayed Saturation
Range
“<50” for levels below 50%; 50 to 100%
Saturation Accuracy
70–100% (0–69% unspecified)
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)
CHARACTERISTIC
DESCRIPTION