beautypg.com

No implied license, Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 71

background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-27

No Implied License

Possession or purchase of the pulse oximeter does not convey any expressed or implied license to

use the pulse oximeter with unauthorized sensors or cables which would, alone or in combination

with this device, fall within the scope of one or more of the patents relating to this device.

INACCURATE PULSE OXIMETER READINGS

The pulsations from intra-aortic balloon support can be additive to the pulse rate on

the oximeter pulse rate display. Verify patient’s pulse rate against the ECG heart rate.

POSSIBLE SKIN INJURY

Prolonged, continuous use of a sensor may cause irritation, blistering, or pressure

necrosis of the skin. Check the sensor site regularly based on patient condition and

type of sensor. Change the sensor site if skin changes occur. Do not use tape to hold

the sensor in place as this may cause inaccurate readings or damage to the sensor or

skin.

POSSIBLE STRANGULATION

Carefully route patient cabling to reduce the possibility of patient entanglement or

strangulation.

CAUTIONS

EQUIPMENT HAZARDS

POSSIBLE EQUIPMENT DAMAGE

To avoid damage to the cable, always hold by the connector rather than the cable,

when connecting or disconnecting either end.

POSSIBLE EQUIPMENT DAMAGE

Do not soak or immerse the sensors or cables in any liquid solution. Do not attempt to

sterilize.

WARNINGS (CONTINUED)