MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-27
No Implied License
Possession or purchase of the pulse oximeter does not convey any expressed or implied license to
use the pulse oximeter with unauthorized sensors or cables which would, alone or in combination
with this device, fall within the scope of one or more of the patents relating to this device.
INACCURATE PULSE OXIMETER READINGS
The pulsations from intra-aortic balloon support can be additive to the pulse rate on
the oximeter pulse rate display. Verify patient’s pulse rate against the ECG heart rate.
POSSIBLE SKIN INJURY
Prolonged, continuous use of a sensor may cause irritation, blistering, or pressure
necrosis of the skin. Check the sensor site regularly based on patient condition and
type of sensor. Change the sensor site if skin changes occur. Do not use tape to hold
the sensor in place as this may cause inaccurate readings or damage to the sensor or
skin.
POSSIBLE STRANGULATION
Carefully route patient cabling to reduce the possibility of patient entanglement or
strangulation.
CAUTIONS
EQUIPMENT HAZARDS
POSSIBLE EQUIPMENT DAMAGE
To avoid damage to the cable, always hold by the connector rather than the cable,
when connecting or disconnecting either end.
POSSIBLE EQUIPMENT DAMAGE
Do not soak or immerse the sensors or cables in any liquid solution. Do not attempt to
sterilize.