Therapy – Physio-Control LIFEPAK 15 User Manual
Page 135
![background image](/manuals/313537/135/background.png)
THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-21
For general troubleshooting tips, see Table 10-2 on page 10-18.
Home Screen is blank but ON
LED is illuminated
Screen not functioning
properly
• Press ANALYZE and follow
voice prompts to treat
patient.
Analysis result is NO SHOCK
ADVISED
and ECG shows a
perfectly flat, isoelectric line.
The Test Load is connected to
therapy cable
• Remove the Test Load and
connect therapy electrodes to
the cable.
Table 5-1 Troubleshooting Tips for AED Mode (Continued)
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION