LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Synchronized Cardioversion Procedure

7. Prepare the patient’s skin and apply therapy electrodes to the patient in the anterior-lateral

position. (See "Therapy Electrode and Standard Paddle Placement" on page 5-4.) If using

standard paddles, apply conductive gel to the paddles and place paddles on the patient’s chest.

8. Press

ENERGY SELECT

or rotate the

SPEED DIAL

to select the desired energy. On the standard (hard)

paddles, rotate the

ENERGY SELECT

dial.

9. Press

CHARGE

. While the defibrillator is charging, a charging bar appears and a ramping tone

sounds, indicating the charging energy level. When the defibrillator is fully charged, the screen

displays available energy.

10.Make certain all personnel, including the operator, stand clear of the patient, bed, stretcher,

and any equipment connected to the patient.

11.Confirm ECG rhythm. Confirm available energy.

12.Press and hold the

(shock) button on the defibrillator until the

ENERGY DELIVERED

message

appears on the screen. For standard paddles, press and hold both

(shock) buttons on the

paddles simultaneously until the

ENERGY DELIVERED

message appears on the screen. Release

buttons. For safety reasons, the

(shock) button on the defibrillator front panel is disabled

when using standard paddles.

Note:

To disarm (cancel a charge), press the

SPEED DIAL

. The defibrillator disarms automatically

if shock buttons are not pressed within 60 seconds, or if you change the energy selection after

charging begins.

13.Observe patient and ECG rhythm. Repeat procedure starting from Step 4, if necessary.