No shock advised, Automated external defibrillation (aed) – Physio-Control LIFEPAK 15 User Manual
Page 126
![background image](/manuals/313537/126/background.png)
5-12
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Automated External Defibrillation (AED)
No Shock Advised
The following prompts occur if no shock is advised:
AED Mode
Push ANALYZE
When the CPR countdown time ends, the
PUSH ANALYZE
prompts occur. These prompts
repeat every 20 seconds until you press
ANALYZE
.
AED Mode
No shock advised
If the AED detects a nonshockable rhythm,
the NO SHOCK ADVISED prompts occur. The
defibrillator does not charge, and no shock
can be delivered.
AED Mode
Start CPR
1:46
After NO SHOCK ADVISED, the START CPR
prompts occur. A countdown timer (min:sec
format) continues for the duration specified in
the CPR TIME 2 setup option.
Note: The CPR metronome automatically
provides audible compression “tocks” and
ventilation prompts or tones only during CPR
intervals. To silence the metronome, press
CPR
. To un-silence the metronome, press CPR
again.
AED Mode
Push ANALYZE
When the CPR countdown time ends, the
PUSH ANALYZE
prompts occur. These prompts
repeat every 20 seconds until you press
ANALYZE
.