Troubleshooting tips, Therapy – Physio-Control LIFEPAK 15 User Manual
Page 143

THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-29
Troubleshooting Tips
Table 5-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Charge time to 360 joules
exceeds 10 seconds
Battery low
• Replace battery with fully
charged battery.
Operating temperature is too
low
• Move patient and device to
warmer environment, if
necessary.
Energy not delivered to patient
when
(shock)
buttons are
pressed
Device is in Sync mode and
QRS complexes are not
detected
• Adjust ECG size for
optimum sensing QRS or
deactivate SYNC if rhythm
VF/VT.
SYNC
accidentally pressed and
rhythm is VF/VT
• Press SYNC to turn off Sync.
• Press
(shock)
buttons.
Device in Sync mode and
(shock)
buttons not pressed
and held until next detected
QRS
• Hold
(shock)
buttons
until discharge occurs or
next detected QRS and
ENERGY DELIVERED
message
appears.
(shock)
buttons pressed
before full charge reached
• Wait for tone and message
indicating full charge.
Standard paddles connected
and
(shock)
button on
defibrillator front panel pressed
• Simultaneously press
(shock)
buttons on standard
paddles to discharge.
Sixty seconds elapsed before
(shock)
buttons were
pressed after full charge.
Energy was internally removed.
• Press
(shock)
buttons
within 60 seconds of full
charge.
Energy selection changed
• Press CHARGE again.
CONNECT CABLE
message
appears
Therapy cable disconnected
during charging
• Reconnect cable and press
CHARGE
again.
Therapy cable damaged
• Replace therapy cable and
perform daily checks per
Operator’s Checklist.
ENERGY FAULT
message appears
(selected and available energy)
Defibrillator out of calibration
• Attempt to transfer energy.
• Contact a qualified service
technician.