General dangers and warnings – Physio-Control LIFEPAK 15 User Manual
Page 16
2-4
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
General Dangers and Warnings
SHOCK OR FIRE HAZARD
Do not immerse any portion of this defibrillator in water or other fluids. Avoid spilling
any fluids on defibrillator or accessories. Spilled liquids may cause the defibrillator
and accessories to perform inaccurately or fail. Do not clean with ketones or other
flammable agents. Do not autoclave or sterilize this defibrillator or accessories unless
otherwise specified.
POSSIBLE FIRE
Use care when operating this device close to oxygen sources (such as bag-valve-mask
devices or ventilator tubing). Turn off gas source or move source away from patient
during defibrillation.
ELECTRICAL INTERFERENCE HAZARDS
POSSIBLE ELECTRICAL INTERFERENCE WITH DEVICE PERFORMANCE
Equipment operating in close proximity may emit strong electromagnetic or radio
frequency interference (RFI), which could affect the performance of this device. If use
of equipment in close proximity is necessary, observe the device to verify normal
operation in the configuration in which the device will be used. RFI may result in
distorted ECG, incorrect ECG lead status, failure to detect a shockable rhythm,
cessation of pacing, or incorrect vital sign measurements. Avoid operating the device
near cauterizers, diathermy equipment, or other portable and mobile RF
communications equipment. Do not rapidly key EMS radios on and off. Refer to
Appendix D for recommended distances of equipment. Contact Physio-Control
Technical Support if assistance is required.
POSSIBLE ELECTRICAL INTERFERENCE
Using cables, electrodes, or accessories not specified for use with this defibrillator
may result in increased emissions or immunity from electromagnetic or radio
frequency interference (RFI) which could affect the performance of this defibrillator or
of equipment in close proximity. Use only parts and accessories specified in these
operating instructions.
POSSIBLE ELECTRICAL INTERFERENCE
This defibrillator may cause electromagnetic interference (EMI) especially during
charge and energy transfers. EMI may affect the performance of equipment operating
in close proximity. Verify the effects of defibrillator discharge on other equipment prior
to using the defibrillator in an emergency situation, if possible.
IMPROPER DEVICE PERFORMANCE HAZARDS
POSSIBLE IMPROPER DEVICE PERFORMANCE
Using other manufacturers’ cables, electrodes, power adapters, or batteries may cause
the device to perform improperly and may invalidate the safety agency certifications.
Use only the accessories that are specified in these operating instructions.
WARNINGS (CONTINUED)