3-10
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Front View
Connectors
Figure 3-7 Connectors for IP Monitoring Configuration
Note:
If your LIFEPAK 15 monitor/defibrillator is configured for temperature monitoring, P1 and P2
are replaced by a single port labeled TEMP. For more information, see Figure 3-8 on page 3-11.
CO
2
SpO2
NIB
P
ECG
P1
P2
CONNECTOR
ACTION
CO2
Connect: Open CO
2
port door, insert FilterLine connector, and turn
clockwise until connector is firmly seated.
Disconnect: Rotate FilterLine connector counterclockwise and pull
connector out.
SpO2/
SpCO/
SpMet
Connect: Align cable connector with SpO
2
port and push in until
connector clicks into place.
Disconnect: Press the gray buttons on each side of the cable
connector simultaneously and pull connector out.
NIBP
Connect: Insert NIBP tubing connector into the NIBP port.
Disconnect: Press the latch on the left side of the port and pull
tubing connector out.
ECG
Connect: Align the green ECG connector with the ECG port;
position the white line on the cable facing left. Insert the cable
connector into the port until the connector is firmly seated.
Disconnect: Pull the ECG connector straight out.
P1/P2
Connect: Align the IP (invasive pressure) cable connector with the
P1 or P2 port; position the gap on the connector facing up. Insert
the cable connector into the port until the connector is firmly
seated.
Disconnect: Grip the connector and pull straight out.