beautypg.com

Connectors, Front view – Physio-Control LIFEPAK 15 User Manual

Page 28

background image

3-10

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Front View

Connectors

Figure 3-7 Connectors for IP Monitoring Configuration

Note:

If your LIFEPAK 15 monitor/defibrillator is configured for temperature monitoring, P1 and P2

are replaced by a single port labeled TEMP. For more information, see Figure 3-8 on page 3-11.

CO

2

SpO2

NIB

P

ECG

P1

P2

CONNECTOR

ACTION

CO2

Connect: Open CO

2

port door, insert FilterLine connector, and turn

clockwise until connector is firmly seated.

Disconnect: Rotate FilterLine connector counterclockwise and pull
connector out.

SpO2/
SpCO/

SpMet

Connect: Align cable connector with SpO

2

port and push in until

connector clicks into place.

Disconnect: Press the gray buttons on each side of the cable
connector simultaneously and pull connector out.

NIBP

Connect: Insert NIBP tubing connector into the NIBP port.

Disconnect: Press the latch on the left side of the port and pull
tubing connector out.

ECG

Connect: Align the green ECG connector with the ECG port;
position the white line on the cable facing left. Insert the cable
connector into the port until the connector is firmly seated.

Disconnect: Pull the ECG connector straight out.

P1/P2

Connect: Align the IP (invasive pressure) cable connector with the
P1 or P2 port; position the gap on the connector facing up. Insert
the cable connector into the port until the connector is firmly
seated.

Disconnect: Grip the connector and pull straight out.