Table a-3 list, A-14, Specifications and performance characteristics – Physio-Control LIFEPAK 15 User Manual
Page 248

A-14
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
Table A-3 Alarm Limits
VITAL SIGN
(VS)
PATIENT
VS VALUE
WIDE LIMITS*
NARROW LIMITS*
LIMITS RANGE
DEFAULT LIMITS**
LOW
HIGH
LOW
HIGH
LOW
HIGH
LOW
HIGH
Heart Rate
(HR)
Pulse Rate
(PR)
<60
-20
+35
-10
+25
30–150 100–250 50
150
60–79
-25
+40
-20
+30
80–104
-30
+40
-30
+30
(bpm)
>105
-35
+45
-25
+25
SpO
2
(%)
>90
-5
+3
-5
+3
50
90–100
85
100
<90
-5
+3
-5
+3
Systolic BP
(mmHg)
<90
-20
+35
-10
+25
30
245
50
200
90–114
-20
+35
-10
+25
115–140
-25
+35
-10
+20
>140
-25
+35
-10
+20
Diastolic BP
(mmHg)
<65
-15
+25
-10
+25
12
210
20
150
65-90
-15
+15
-15
+10
>90
-15
+15
-15
+10
EtCO
2
(mmHg/%)
>40/5.3
-10/-1.3 +15/+2.0 -10/-1.3 +15/+2.0 5/0.7
70/9.2
15/2.0
50/6.6
40/5.3
-10/-1.3 +15/+2.0 -10/-1.3 +15/+2.0
Inspired CO
2
(mmHg/%)
—
n/a
+5/+0.7
n/a
+3/+0.4
n/a
0/0–10/
1.3
—
8/1.1
Respiration
Rate (RR)
<15
-8
+8
-4
+4
5-15
10–60
5
30
15
-15
+15
-8
+8
Systolic PA
(mmHg)
<15
-6
+12
-4
+6
10
100
10
40
15–35
-8
+16
-6
+8
>35
-12
+16
-8
+10
Diastolic PA
(mmHg)
<5
-4
+12
-4
+8
0
50
0
18
5–13
-4
+16
-6
+6
>13
-6
+16
-6
+6
CVP (mmHg)
9
-10
+10
-5
+5
0
25
0
15
ICP, LAP
(mmHg)
<15
-6
+6
-4
+4
0
40
0
18
15
-6
+8
-4
+6
Temperature
(°C)
-3
+3
-1
+1
31
41
35
39
*Numbers are ± from patient’s VS value when the alarms are set.
**Default limits are established when alarms are set up to be ON.