THERAPY

5

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-15

AED/Passcode—A passcode screen appears:

Restricted—A

MANUAL MODE DISABLED

message appears, an alert tone sounds, and the

LIFEPAK 15 monitor/defibrillator returns to AED mode.

It is important that all users of the LIFEPAK 15 monitor/defibrillator be thoroughly familiar with the

monitor/defibrillator settings and operation before use.

Special AED Setup Options

The following descriptions of AED prompts (voice and text) explain special setup options.

Initial CPR - CPR First

When the

INITIAL CPR

option is set to

CPR FIRST

, you are prompted to

START CPR

immediately after

the AED is turned on, and before an analysis.

Manual Mode

Enter Manual Mode?

No

Push Speed Dial to confirm

Select YES to enter Manual mode.

Yes

Manual Mode

Enter Manual Mode passcode

Rotate and press the SPEED DIAL to enter the
passcode.

The code changes to dots to protect the
passcode, and the defibrillator enters Manual
mode.

You have three opportunities to enter the
correct password. After an incorrect attempt,
the message INCORRECT--TRY AGAIN appears.
After three incorrect attempts, the message
ACCESS DENIED

appears, and the defibrillator

returns to AED mode.