Shock advised, Automated external defibrillation (aed) – Physio-Control LIFEPAK 15 User Manual
Page 124

5-10
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Automated External Defibrillation (AED)
6. Press
ANALYZE
to initiate the analysis. Stop CPR.
7. Continue to follow the screen messages and voice prompts provided by the AED.
Shock Advised
The following prompts occur when shock is advised:
WARNING
POSSIBLE MISINTERPRETATION OF DATA
Do not move the AED during analysis. Moving the AED during analysis may affect the
ECG signal resulting in an inappropriate
SHOCK
or
NO SHOCK ADVISED
decision. Do not
touch the patient or the AED during analysis.
The ANALYZING NOW—STAND CLEAR prompts
occur. The SAS analyzes the patient’s ECG in
approximately 6 to 9 seconds and advises
either SHOCK ADVISED
or NO SHOCK ADVISED.
AED Mode
Analyzing now -- Stand clear
If the AED detects a shockable rhythm, the
SHOCK ADVISED
prompts occur. Charging to
the joule setting for Shock #1 begins. A
charging bar appears and a ramping tone
sounds.
AED Mode
SHOCK ADVISED!
Charging to 200J
200J
To cancel, push Speed Dial