LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Specifications and Performance Characteristics

Waveform Shape and Measured Parameters

Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)

CHARACTERISTIC

DESCRIPTION

T1

T2

Patient

Impedance (

)

25
50
75

100
125
150
175

Phase 1 Duration (ms)

Phase 2 Duration (ms)

Tilt (%)

Min
5.1
6.8
7.6
8.7
9.5

10.1
10.6

Max

6.0
7.9
9.4

10.6
11.2
11.9
12.5

Min
3.2
4.4
4.9
5.6
6.2
6.6
6.9

Max

4.2
5.5
6.5
7.3
7.7
8.2
8.6

Min

69.9
57.0
49.3
43.0
39.0
36.8
33.8

Max

85.2
74.7
67.6
62.2
56.6
52.6
49.3

I1

I2

I3

I4

Biphasic Waveform