Appendix – Physio-Control LIFEPAK 15 User Manual

Page 257

background image

APPENDIX

B

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

B-3

CPR: YOUTH-AIRWAY X:Y

An option for CPR metronome. The patient is a youth (younger
than the age of puberty) for whom an advanced airway has been
established. The specified C:V ratio will be used.

CPR: YOUTH-NO AIRWAY X:Y

An option for CPR metronome. The patient is a youth (younger
than the age of puberty) for whom an advanced airway has not
been established. The specified C:V ratio will be used.

CURRENT FAULT

The comparison between delivered and selected pacing current
is out of tolerance.

DEMAND

Pacemaker is in Demand mode.

DEMO MODE

The device is in Demo mode and simulated patient data is
displayed.

DISARMING...

The energy charge is being removed internally.

ECG CABLE OFF

The device is printing and the ECG cable is removed.

ECG LEADS OFF

Multiple ECG electrodes are disconnected.

ENDING DEVICE SEARCH

The request for finding a Bluetooth device was stopped.

ENERGY DELIVERED

Energy transfer is complete.

ENERGY FAULT

The comparison between stored and selected energy is out of
tolerance.

ENTER MANUAL MODE?

One of the Manual mode access buttons was pressed and the
confirmation screen is set up to appear.

EXCESSIVE NOISE - 12-LEAD
CANCELLED

Noise is detected for longer than 30 seconds that is too great to
record a 12-lead ECG report.

IF NO PULSE, PUSH ANALYZE

Follows a CPR interval, if a PULSE CHECK setup option other than
NEVER

is selected.

IF NO PULSE, START CPR

Follows delivery of a shock or NO SHOCK ADVISED prompt, if a
PULSE CHECK

setup option other than NEVER is selected.

IF YOU WITNESSED THE ARREST,
PUSH ANALYZE

Initial CPR message that follows START CPR prompt, to remind
user to deliver a shock immediately if the user witnessed the
arrest.

LA LEADS OFF

ECG electrode “LA” is disconnected.

LAST CONNECTED TO

When Bluetooth connectivity is installed and this device
previously connected to a target device, the name of the target
device appears after this message.

LL LEADS OFF

ECG electrode “LL” is disconnected.

LOST BLUETOOTH CONNECTION

Communication with Bluetooth device has been interrupted.

LOST DIRECT CONNECTION

Communication via direct connection has been interrupted.

Table B-1 Summary of Screen Messages (Continued)

MESSAGE

DESCRIPTION