Monitoring – Physio-Control LIFEPAK 15 User Manual
Page 109
MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-65
Temperature area of home
screen is blank
Initial temperature not
automatically displayed
until device detects
temperature between 24.8°
and 45.2°C (76.6° and
113.4°F)
• Allow up to 3 minutes for probe to
equilibrate.
• Check that probe is positioned
properly.
Temperature probe not
detected by device
• Check connections between probe,
adapter cable, and device.
• Check that the sensor is approved
for use with the LIFEPAK 15
monitor/defibrillator.
• Contact qualified service personnel.
Table 4-11 Troubleshooting Tips for Temperature Monitoring (Continued)
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION