MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-65

Temperature area of home
screen is blank

Initial temperature not
automatically displayed
until device detects
temperature between 24.8°
and 45.2°C (76.6° and
113.4°F)

• Allow up to 3 minutes for probe to

equilibrate.

• Check that probe is positioned

properly.

Temperature probe not
detected by device

• Check connections between probe,

adapter cable, and device.

• Check that the sensor is approved

for use with the LIFEPAK 15
monitor/defibrillator.

• Contact qualified service personnel.

Table 4-11 Troubleshooting Tips for Temperature Monitoring (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION