Troubleshooting tips, Noninvasive pacing – Physio-Control LIFEPAK 15 User Manual
Page 150

5-36
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Noninvasive Pacing
Troubleshooting Tips
Table 5-4 Troubleshooting Tips for Noninvasive Pacing
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Device does not function
when PACER is pressed
Power off
• Check if power is ON.
Low battery
• Replace battery with fully
charged battery.
PACER
LED is on, but
CURRENT (mA)
will not
increase
Therapy electrodes off
• Check for message displayed.
• Inspect therapy cable and
electrode connections.
PACER
LED on,
CURRENT (mA) >0,
but
pace markers absent (not
pacing)
Pacing rate set below patient's
intrinsic rate
• Increase PPM.
Pacer oversensing (ECG artifact,
ECG size too high)
• Establish clean ECG;
decrease ECG size.
• Select nondemand pacing.
Monitor screen displays
distortion while pacing
ECG electrodes not optimally
placed with respect to pacing
electrodes
• Reposition electrodes away
from pacing electrodes.
• Select another lead (I, II,
or III).
Pacing stops spontaneously
PACER
pressed off
• Press PACER and increase the
current.
Internal error detected. Service
message indicates an internal
failure.
• Check for service indicator.
• Cycle power and start pacing
again.
• Obtain service by a qualified
service technician.
Therapy electrode off
• Check for message. Check
pacing cable and electrode
connections.
CHARGE
pressed
• Press PACER and increase
current, if pacing desired.
Otherwise, proceed with
defibrillation.
Radio frequency interference
• Move radio equipment away
from pacemaker.