Troubleshooting tips, Data transmission – Physio-Control LIFEPAK 15 User Manual
Page 193

DATA TRANSMISSION
8
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
8-15
Troubleshooting Tips
Table 8-3 Troubleshooting Tips for Data Transmission
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Bluetooth icon on
LIFEPAK 15 monitor has
red X across it
WIRELESS
is set to OFF in the
Bluetooth Setup menu
• Set WIRELESS to ON.
• If red X remains, Bluetooth
module in LIFEPAK 15 monitor
may be faulty. Contact qualified
service representative.
WIRELESS
is set to OFF in the setup
options, so the WIRELESS default
is OFF each time the LIFEPAK 15
monitor is turned on
• Change WIRELESS setup option.
See LIFEPAK 15 Monitor/
Defibrillator Setup Options
provided with your device.
• If red X remains, Bluetooth
module in LIFEPAK 15 monitor
may be faulty. Contact qualified
service representative.
Bluetooth module in LIFEPAK 15
monitor may be faulty
• Contact qualified service
representative.
Bluetooth LED is not
illuminated
Target device is off or cannot
communicate with the
LIFEPAK 15 monitor
• Confirm that target device is on
and discoverable.
• See the operating instructions for
your target device.
Bluetooth module in LIFEPAK 15
monitor may be faulty
• If other troubleshooting is
unsuccessful, contact qualified
service representative.
LIFEPAK 15 monitor
does not automatically
connect to last connected
device
Target device is off or cannot
communicate with the LIFEPAK
15 monitor
• Confirm that target device is on
and discoverable.
Last connection to target device
may have occurred when the
LIFEPAK 15 monitor was in a
different mode
• Confirm that OUTPUT PORT is set to
BLUETOOTH WIRELESS
.
• Select FIND DEVICES and establish
a new connection.
Device does not connect
to last connected device
after WIRELESS is set to
ON
Bluetooth menu is displayed,
which prevents discovery of
devices
• Press HOME SCREEN to exit menu
and allow LIFEPAK 15 monitor to
find last connected device.
UNABLE TO CONNECT
message appears
LIFEPAK 15 monitor cannot
establish wireless connection.
Target device may not have the
necessary software application or
cannot accept data.
• Verify target device is ready to
receive transmissions.
• Attempt to retransmit.