Table a-2 lists, Appendix, A-13 – Physio-Control LIFEPAK 15 User Manual
Page 247
![background image](/manuals/313537/247/background.png)
APPENDIX
A
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
A-13
Bump
1000 bumps at 15 g with pulse duration of 6 msec
Impact, Non-operating
EN 60601-1 0.5 + 0.05 joule impact
UL 60601-1 6.78 Nm impact with 2-inch diameter steel ball
Meets IEC62262 protection level IK 04
EMC
EN 60601-1-2:2001 Medical Equipment - General Requirements for
Safety - Collateral Standard: Electromagnetic Compatibility -
Requirements and Tests
EN 60601-2-4:2003: (Clause 36) Particular Requirements for the
Safety of Cardiac Defibrillators and Cardiac Defibrillator-Monitors
Cleaning
Cleaning 20 times with the following: Quaternary ammonium, isopropyl
alcohol, hydrogen peroxide
Chemical Resistance
60 hour exposure to specified chemicals:
Betadine (10% Povidone-Iodine solution)
Coffee, Cola
Dextrose (5% Glucose solution)
Electrode Gel/Paste (98% water, 2% Carbopol 940)
HCL (0.5% solution, pH=1)
Isopropyl Alcohol
NaCl solution (0.9% solution)
Cosmetic discoloration of the paddle well shorting bar shall be allowed
following exposure to HCL (0.5% solution).
Table A-2 Battery Specifications
CHARACTERISTIC
DESCRIPTION
Battery Type
Lithium-ion
Weight
< 0.60 kg (1.3 lb)
Charge Time (with fully depleted battery)
4 hours and 15 minutes (typical)
Battery indicators
Each battery has a fuel gauge that indicates its
approximate charge. A fuel gauge that shows two or fewer
LEDs after a charge cycle indicates that the battery
should be replaced.
Charging Temperature Range
5° to 45°C (41° to 113°F)
Operating Temperature Range
0° to 45°C (32° to 113°F)
Short Term (<1 week) Storage Temperature
Range
-20° to 60°C (-4° to 140°F)
Long Term (>1 week) Storage Temperature
Range
20° to 25°C (68° to 77°F)
Operating and Storage Humidity Range
5 to 95% relative humidity, non-condensing
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)