beautypg.com

Aed warnings, Aed mode, Automated external defibrillation (aed) – Physio-Control LIFEPAK 15 User Manual

Page 122: Aed warnings aed mode

background image

5-8

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Automated External Defibrillation (AED)

AED Warnings

AED Mode

The LIFEPAK 15 monitor/defibrillator is set up to operate in Manual mode when it is turned on

(factory default setting). The device can be set up to power on in AED mode by changing the Setup

Options. The factory default settings for AED mode are identified in Table A-5 on page A-16. The

energy settings and other AED setup options can be changed according to medical protocol. For

more information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your

device.

The ECG is continuously displayed in AED mode; however, access to other functions such as

OPTIONS

is not allowed in AED mode. The CPR metronome automatically sounds during CPR times,

but it can only be silenced and un-silenced in AED mode. For more information, see "CPR Time and

Metronome" on page 5-14.

You can exit AED mode’s prompted protocol and enter Advisory Monitoring or Manual Mode. For

more information about Advisory Monitoring, see "Advisory Monitoring" on page 5-18. Access to

Manual mode may be direct, require confirmation or a passcode, or not allowed, depending on how

your defibrillator is set up. It is important to be thoroughly familiar with your monitor/defibrillator

settings and operation before use.

WARNINGS

MISINTERPRETATION OF DATA HAZARDS

POSSIBLE MISINTERPRETATION OF DATA

Do not analyze in a moving vehicle. Motion artifact may affect the ECG signal resulting

in an inappropriate

SHOCK

or

NO SHOCK ADVISED

message. Motion detection may delay

analysis. Stop vehicle and stand clear of patient during analysis.

POSSIBLE ECG MISINTERPRETATION

Do not place therapy electrodes in the anterior-posterior position when operating this

defibrillator in AED mode. A

SHOCK

or

NO SHOCK

decision may be inappropriately

advised. The shock advisory algorithm requires the electrodes to be placed in the

anterior-lateral (Lead II) position.

PEDIATRIC PATIENT SAFETY RISK

In AED mode, this defibrillator is not designed or tested to interpret pediatric rhythms

or administer energy at pediatric joule settings for children under eight years old.