Therapy – Physio-Control LIFEPAK 15 User Manual
Page 151
![background image](/manuals/313537/151/background.png)
THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-37
No muscle response to
pacing
Patient’s heart rate may be
greater than noninvasive pacer
ppm
• No action needed.
The Test Load is connected to
therapy cable
• Remove the Test Load and
connect therapy electrodes to
cable.
Patient muscle response is
variable and depends on patient
condition. Muscular response to
pacing is not a reliable indicator
of current delivered.
• No action needed.
Capture does not occur with
pacing stimulus
Current (mA) set too low
• Increase pacing current.
(Administer sedation or
analgesia as needed.)
CONNECT CABLE
or PACING
STOPPED
message appears
Therapy cable damaged
• Replace therapy cable and
perform daily checks per
Operator’s Checklist.
CONNECT ELECTRODES
message appears
Pacing cable or electrode
disconnected
• Reconnect and set current.
Electrodes not adhering to skin
• Prepare skin.
Therapy cable damaged
• Replace therapy cable and
perform daily checks per
Operator’s Checklist.
Electrodes outdated
• Replace electrodes and set
current.
PACING IN PROGRESS
message
appears
CPR
pressed
• Press PACER to stop pacing, if
appropriate, and then press
CPR
.
Pacing stops spontaneously
and PACER FAULT message
appears
Internal error detected
• Cycle power and start pacing
again.
• Obtain service by a qualified
service technician.
Intrinsic QRS complexes not
sensed when pacing
ECG size too low
• Increase ECG size or select
another lead.
Intrinsic QRS complexes are
occurring during pacemaker's
refractory period
• Adjust PPM.
Table 5-4 Troubleshooting Tips for Noninvasive Pacing (Continued)
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION