General maintenance, Maintenance and service, Cleaning – Physio-Control LIFEPAK 15 User Manual
Page 204
9-8
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
General Maintenance
General Maintenance
Maintenance and Service
The power adapter contains no serviceable parts. If the power adapter does not function correctly,
contact your local Physio-Control representative for assistance.
Cleaning
To clean the power adapter:
1. Unplug the power adapter, if it is connected to an auxiliary power source.
2. Clean the power adapter, power cord, and cables with a damp sponge or cloth. Use only the
cleaning agents listed below:
• Quaternary ammonium compounds
• Isopropyl alcohol
• Peracetic (peroxide) acid solutions
• Sodium dichloroisocyanurate (NaDCC)
• Chlorine bleach (1:10 dilution)
Note:
Carefully clean the connector ports. Do not allow cleaning fluids to penetrate the exterior
surfaces of the device.
WARNING
POSSIBLE ELECTRICAL SHOCK
Unplug the power adapter from the power source before cleaning.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Do not clean any part of the power adapter or its accessories with phenolic
compounds. Do not use abrasive or flammable cleaning agents. Do not attempt to
sterilize this device or any accessories unless otherwise specified in accessory
operating instructions.