THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-11
When charging is complete, the available
energy is displayed.
The STAND CLEAR, PUSH SHOCK BUTTON! (
)
message occurs, followed by a “Shock ready”
tone.
Clear everyone away from touching the
patient, bed, or any equipment that is
connected to the patient.
Press
(shock) to deliver energy to the
patient.
AED Mode
200J available
Push SHOCK button!
To cancel, push Speed Dial
AED Mode
Energy delivered
When the
(shock) button is pressed, the
ENERGY DELIVERED
message occurs indicating
that the energy transfer was completed.
AED Mode
Disarming...
Note: If you do not press the
(shock)
button within 60 seconds, or the SPEED DIAL is
pressed to cancel charging, the defibrillator
disarms and the DISARMING message appears.
AED Mode
Start CPR
1:46
After a shock is delivered, the START CPR
prompts occur. A countdown timer (min:sec
format) continues for the duration specified
in the CPR TIME 1 setup option.
Note: The CPR metronome automatically
provides audible compression “tocks” and
ventilation prompts or tones only during CPR
intervals at a ratio of 30:2. To silence the
metronome, press CPR. To un-silence the
metronome, press CPR again.