Waveform events, Figure 7-2, Patient records and reports – Physio-Control LIFEPAK 15 User Manual
Page 170
7-6
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Patient Records and Reports
Waveform Events
In addition to being documented in the Event Log, therapy and other selected events also
capture waveform data that are printed with the long and medium CODE SUMMARY report. The
waveform events and the characteristics of waveform data are described in Table 7-3.
Figure 7-2 Possible Event Log Entries
Monitoring
• Check patient
• Initial rhythm
• Replace battery
• 12-lead
• NIBP
• Alarm events
• IP label change
• Vital signs
• 5-wire on/off
• SpCO/SpMet Advisory
AED
• Connect electrodes
• Motion
• Analysis
• Analysis stopped
• Shock advised
• No shock advised
CPR Metronome
• On/Off
• Age-Airway changed
Defibrillation
• Manual mode
• Charge removed
• Shock X, XXXJ
• Shock X, Abnormal
Operator Initiated
• Event
• Alarms on/off
• Print
• VF/VT alarm on/off
• Sync on/off
• Snapshot
• Internal pacer detection on/off
Pacing
• Started
• Set
• Changed
• Stopped
• Paused
Transmission
• Transmission complete
• Transmission failed
• Transmission cancelled
Memory Status
• Out of waveform memory (memory low)
• Out of event memory (memory full)
Table 7-3 Waveform Events
EVENT NAME
WAVEFORM DATA (WHEN CAPTURED)
INITIAL RHYTHM
8 seconds after leads on
CHECK PATIENT
8 seconds prior to alert
SHOCK or NO SHOCK ADVISED
2-3 segments of analyzed ECG. Each segment is 2.7 seconds
ANALYSIS X STOPPED
8 seconds of data prior to cessation of analysis