Specifications and performance characteristics, Appendix – Physio-Control LIFEPAK 15 User Manual
Page 235
APPENDIX
A
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
A-1
Specifications and Performance Characteristics
Table A-1 lists the LIFEPAK 15 monitor/defibrillator specifications for the device.
Table A-2 lists the specifications for the LIFEPAK 15 monitor/defibrillator batteries.
Table A-3 lists the high and low limits for alarms when either the wide or narrow alarm setting is
selected on the LIFEPAK 15 monitor/defibrillator.
Table A-4 lists the alarm performance characteristics.
Table A-5 lists the factory default settings for the LIFEPAK 15 monitor/defibrillator setup options.
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications
CHARACTERISTIC
DESCRIPTION
All specifications are at 20°C unless otherwise stated.
GENERAL
Classification
Monitor/defibrillator—Battery powered and Class II (per EN 60601-1)
Applied parts—ECG, Invasive Pressure and Temperature are Type CF
patient connections. Therapy, CO2, SpO2, and NIBP are Type BF patient
connections (per EN 60601-1).
Modes
AED mode—for automated ECG analysis and a prompted treatment
protocol for patients in cardiac arrest.
Manual mode—for performing manual defibrillation, synchronized
cardioversion, noninvasive pacing, and ECG and vital sign monitoring.
Archive mode—for accessing stored patient information.
Setup mode—for changing default settings of the operating functions.
Service mode—for authorized personnel to perform diagnostic tests and
calibrations.
Demo mode—for simulated waveforms and trend graphs for
demonstration purposes.
Self-test
When powered on, the device performs a self-test to check internal
electrical components and circuitry. A service indicator is illuminated if
an error is detected.
The device also performs an auto test daily. Results are printed and
stored in the device log. Auto test results can be transmitted. See the
LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your
device for more information.
Continuous Patient
Surveillance System
(CPSS)
In Advisory Monitoring, CPSS monitors the patient ECG, via QUIK-
COMBO
®
electrodes or Lead II, for a potentially shockable rhythm.
Voice Prompts
Manual mode: Used for selected prompts (selectable ON/OFF)
AED mode: Used for entire AED protocol