Dc power adapter operation, Using the power adapter – Physio-Control LIFEPAK 15 User Manual
Page 202

9-6
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Using the Power Adapter
DC Power Adapter Operation
To use the DC Power Adapter:
1. Connect the DC power cable to the power adapter and a 12 Vdc power source.
2. Verify that the green LED strip illuminates.
3. Connect the power adapter output cable to the power adapter.
4. Connect the green end of the power adapter output cable to the auxiliary power connector on the
back of the monitor/defibrillator.
5. Verify that the
AUXILIARY POWER
LED on the defibrillator is illuminated.
6. If at least one battery is installed in the device, verify that the
BATTERY CHARGING
indicator is
illuminated or flashing. Indicator behaviors are shown in Table 9-2.
7. Press the defibrillator
ON
button.
Table 9-2 Battery Charging Indicator Behaviors
INDICATOR
DESCRIPTION
Steady green
Installed batteries are fully charged.
Flashing green
One or both installed batteries are being charged.
Off
No batteries are installed, or a battery is unable to be charged.