LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Specifications and Performance Characteristics

DEFIBRILLATOR

Charge Time (per EN 60601-2-4)

AC Operation Only:

Maximum Time from Charge to Shock Ready (Manual Mode):

Voltage

Charge Time

90-240 Vac (Nominal):

360 J within 10

seconds

Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):

Voltage

Charge Time

90-240

Vac (Nominal)

360 J within 30 seconds

Maximum Time from Power-on to Shock Ready (Manual Mode):

Voltage

Charge Time

90-240 Vac (Nominal)

360 J within 25 seconds

Maximum Time from Power-on to Shock Ready (AED Mode):

Voltage

Charge Time

90-240 Vac (Nominal)

360 J within 40 seconds

DC Operation Only:

Maximum Time from Charge to Shock Ready (Manual Mode):

Voltage

Charge Time

11-17.6 Vdc (Nominal)

360 J within 10 seconds

Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):

Voltage

Charge Time

11-17.6 Vdc (Nominal)

360 J within 30 seconds

Maximum Time from Power-on to Shock Ready (Manual Mode):

Voltage

Charge Time

11-17.6 Vdc (Nominal)

360 J within 25 seconds

Maximum Time from Power-on to Shock Ready (AED Mode):

Voltage

Charge Time

11-17.6 Vdc (Nominal)

360 J within 40 seconds

Battery Operation Only:

Maximum Time from Charge to Shock Ready (Manual Mode):

Battery Status

Charge Time

Fully charged

200 J within 7 seconds, nominal

Fully charged, followed by 15
full-energy shocks

360 J within 10 seconds

Fully charged

360 J within 10

seconds

Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)