Data transmission – Physio-Control LIFEPAK 15 User Manual
Page 195
![background image](/manuals/313537/195/background.png)
DATA TRANSMISSION
8
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
8-17
Unable to transmit using
a gateway device that has
a functioning direct
connection or Bluetooth
connection
Transmission sites are not set up
in LIFEPAK 15 monitor
• Define transmission sites. Each
site name must exactly match the
name of the target device. See
LIFEPAK 15 Monitor/Defibrillator
Setup Options provided with your
device.
Transmission site names in
LIFENET System do not match
site names in LIFEPAK 15
monitor
• Check site names in LIFENET
System.
Cellular communication is not
working between the gateway and
transmission sites
• Use alternate method to
communicate patient data.
UNABLE TO TRANSMIT
message appears
The LIFEPAK 15 monitor cannot
connect to the device name
selected
• Verify target device is ready to
receive transmissions.
• Verify target device setup.
• Attempt to retransmit.
The output port on the
LIFEPAK 15 monitor is not
configured for the transmission
method you are using
• Make sure the transmission site
OUTPUT PORT
is configured for the
type of transmission you are
attempting.
• Attempt to retransmit.
Target device unable to connect or
unable to connect within timeout
interval
• Verify target device is ready to
receive transmissions.
• Verify target device setup.
• Attempt to retransmit.
The target device requires you to
“accept” incoming
communications
• Check your target device for a
required acknowledgment to
connect.
• Enter passcode, when prompted.
• Set to “Always allow” if possible.
• Attempt to retransmit.
Direct connection was disrupted
• Verify cable connections.
• Attempt to retransmit.
TRANSMISSION FAILED
message appears
Computer application program is
not ready or is not available to
receive transmission
• Verify target device is running
necessary software.
• Attempt to retransmit.
Table 8-3 Troubleshooting Tips for Data Transmission (Continued)
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION