Therapy – Physio-Control LIFEPAK 15 User Manual
Page 145
THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-31
ABNORMAL ENERGY DELIVERY
message appears and Shock XJ
Abnormal annotated on printout
Open air discharge with
standard paddles
• Press paddles firmly on
patient’s chest when
discharging.
Standard paddles placed face-
to-face when
(shock)
button
pressed
• Perform test discharges per
Operator’s Checklist.
Patient impedance is out of
range
• Increase energy or repeat
shocks as needed.
• Consider replacing
disposable therapy
electrodes with new ones.
Internal fault occurred
• Repeat shock.
• Perform CPR and obtain
another defibrillator, if
necessary.
CONNECT ELECTRODES
message
appears
Therapy electrodes are not
connected to the therapy cable
• Check for electrode
connection.
Electrodes do not adhere
properly to the patient
• Press electrodes firmly on
patient’s skin.
• Clean, shave, and dry the
patient’s skin as
recommended.
• Apply new electrodes.
Electrodes are dry, damaged, or
out of date
• Apply new electrodes.
Therapy cable damaged
• Replace therapy cable and
perform daily checks per
Operator’s Checklist.
REPLACE BATTERY
prompt occurs
Both batteries are very low
• Replace one or both
batteries immediately.
• Connect to auxiliary power
using approved power
adapter.
CPR time shown (minutes/
seconds) is different than
expected
Metronome is on
• None. The metronome
adjusts the CPR time to
ensure CPR cycle ends with
compressions.
Incorrect setup option selected
• Change CPR time setup
option. See LIFEPAK 15
Monitor/Defibrillator Setup
Options provided with your
device
.
Table 5-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION