APPENDIX

A

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

A-5

Adults/Pediatrics

±2 digits (during no motion conditions)
±3 digits (during motion conditions)

Dynamic signal strength bar graph

Pulse tone as SpO

2

pulsations are detected

SpO

2

Update Averaging

Rate

User selectable: 4, 8, 12 or 16 seconds

SpO

2

Sensitivity

User selectable: Normal, High

SpO

2

Measurement

Functional SpO

2

values are displayed and stored

Pulse Rate Range

25 to 240 bpm

Pulse Rate Accuracy

Adults/Pediatrics

±3 digits (during no motion conditions)
±5 digits (during motion conditions)

Optional SpO

2

waveform display with autogain control

SpCO™

SpCO Concentration
Display Range

0 to 40%

SpCO Accuracy

±3 digits

SpMet™

SpMet Saturation Range 0 to 15.0%

SpMet Display
Resolution

0.1% up to 10%

SpMet Accuracy

±1 digit

NIBP

Blood Pressure

Systolic Pressure Range: 30 to 255 mmHg

Diastolic Pressure Range: 15 to 220 mmHg

Mean Arterial Pressure Range: 20 to 235 mmHg

Units: mmHg

Blood Pressure Accuracy: ±5 mmHg

Blood Pressure Measurement Time: 20 seconds, typical (excluding cuff
inflation time)

Pulse Rate

Pulse Rate Range: 30 to 240 pulses per minute
Pulse Rate Accuracy: ±2 pulses per minute or ±2%, whichever is greater

Operation Features

Initial Cuff Pressure: User selectable, 80 to 180 mmHg
Automatic Measurement Time Interval: User selectable

Automatic Cuff Deflation Excessive Pressure: If cuff pressure exceeds 290 mmHg

Excessive Time: If measurement time exceeds 120 seconds

CO

2

CO

2

Range

0 to 99 mmHg (0 to 13.2 kPa)
Units: mmHg, %, or kPa

Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)