Appendix – Physio-Control LIFEPAK 15 User Manual
Page 239

APPENDIX
A
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
A-5
Adults/Pediatrics
±2 digits (during no motion conditions)
±3 digits (during motion conditions)
Dynamic signal strength bar graph
Pulse tone as SpO
2
pulsations are detected
SpO
2
Update Averaging
Rate
User selectable: 4, 8, 12 or 16 seconds
SpO
2
Sensitivity
User selectable: Normal, High
SpO
2
Measurement
Functional SpO
2
values are displayed and stored
Pulse Rate Range
25 to 240 bpm
Pulse Rate Accuracy
Adults/Pediatrics
±3 digits (during no motion conditions)
±5 digits (during motion conditions)
Optional SpO
2
waveform display with autogain control
SpCO™
SpCO Concentration
Display Range
0 to 40%
SpCO Accuracy
±3 digits
SpMet™
SpMet Saturation Range 0 to 15.0%
SpMet Display
Resolution
0.1% up to 10%
SpMet Accuracy
±1 digit
NIBP
Blood Pressure
Systolic Pressure Range: 30 to 255 mmHg
Diastolic Pressure Range: 15 to 220 mmHg
Mean Arterial Pressure Range: 20 to 235 mmHg
Units: mmHg
Blood Pressure Accuracy: ±5 mmHg
Blood Pressure Measurement Time: 20 seconds, typical (excluding cuff
inflation time)
Pulse Rate
Pulse Rate Range: 30 to 240 pulses per minute
Pulse Rate Accuracy: ±2 pulses per minute or ±2%, whichever is greater
Operation Features
Initial Cuff Pressure: User selectable, 80 to 180 mmHg
Automatic Measurement Time Interval: User selectable
Automatic Cuff Deflation Excessive Pressure: If cuff pressure exceeds 290 mmHg
Excessive Time: If measurement time exceeds 120 seconds
CO
2
CO
2
Range
0 to 99 mmHg (0 to 13.2 kPa)
Units: mmHg, %, or kPa
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)
CHARACTERISTIC
DESCRIPTION