Initialization of calibration – KEYENCE LS-7600 User Manual
Page 90

5-28
Chapter 5 Function Settings
5
5
Press the [ENT] key.
The input value is entered and the screen returns to the AREA selection status of
calibration settings.
Note:
If the calibration conditions are not satisfied, ERR1 will be displayed. Press
the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the
calibration settings again.
Tips
When repeating the one-point calibration, the previous settings will appear on
the screen, which will be different from those in the above setting example.
Initialization of Calibration
Take the following procedure to initialize the calibration values you are setting or you have
already set (i.e., to return the values to the ones set by default).
In this case, the span value is by default set to 1 and the shift value is by default set to 0.
Note:
• It is possible to initialize the calibration value of each area independently.
• All calibration values set by default will be logic calibration values.
1
Press the [ENT] key when the cursor is at RESET.
The cursor moves to CANCEL.
Initializing AREA1 Calibration Value
(1) Press the [Down] key and move
the cursor to RESET.
(2) Press the [ENT] key.
The value is initialized and the screen
returns to the AREA selection status
of calibration settings.
Canceling Initialization of AREA1
Calibration Value
(1) Press the [Up or Down] key and
move the cursor to
CANCEL.
(2) Press the [ENT] key.
The initialization is canceled and the
screen returns to the AREA selection
status of calibration settings.
AREA1
RESET
CANCEL
RESET
AREA1
RESET
CANCEL
RESET