Setting method – KEYENCE LS-7600 User Manual
Page 107

5-45
Chapter 5 Function Settings
5
The following table shows the relationship of scaling values along with their output
ranges and resolutions.
Scaling value [
μ
m/V]
Output range
Resolution (
❈
)
1
±
10
μ
m
0.01
μ
m
2
±
20
μ
m
0.01
μ
m
5
±
50
μ
m
0.02
μ
m
10
±
100
μ
m
0.04
μ
m
20
±
200
μ
m
0.08
μ
m
50
±
500
μ
m
0.2
μ
m
100
±
1mm
0.4
μ
m
200
±
2mm
0.8
μ
m
500
±
5mm
2
μ
m
1000
±
10mm
4
μ
m
2000
±
20mm
8
μ
m
5000
±
50mm
20
μ
m
10000
±
100mm
40
μ
m
❈If the minimum display unit selected is larger than the resolution selected, the mini-
mum display unit will be used as applicable resolution.
■
Setting Method
1
Press the [Up or Down] key and move the cursor to A-UNIT.
2
Press the [ENT] key.
The cursor moves to setting item.
3
Press the [Up or Down] key and
select the value.
4
Press the [ENT] key.
The input item is entered and the cursor
returns to A-Unit.
●
Tolerance Mode with Threshold
Set the measurement values to be output at
10 V and –10 V respectively so that the
scaling value will be decided. The voltage
can be changed while the result waiting
status is “---.-----.” (See Page 5-47)
Measured value
0V
+10.5V
+10V
Voltage
–10V
0
30
–10.5V
(Positive value
over range)
(Negative value
below range)
+10V :30mm
–10V : 0mm
OUT1
A-UNIT
1000um/V
A-SET
OFF
A-CONT +10.5V