KEYENCE LS-7600 User Manual
Page 75

5-13
Chapter 5 Function Settings
5
P (Pitch) Mode
The LS-7600 Series measures objects based on the middle point between the edges of
the objects in this mode.
Bottom
Top
1E
1P
2P
3P
4P
5P
2E
3E
4E
5E
6E
Target center pitch
Edge 0
Edge 0 refers to a method of measurement, in which the center of the optical axis is
used as an origin. In this case, set the edge number to 0.
Use this method to position targets or measure the movement or deviation of targets.
• Use edge 0 method if higher repetitive accuracy is required from the meas-
urement of objects.
• If DUAL is selected in the application function (page 5-68), when no meas-
uring target is in the beam axis, the number of edges becomes “0”.
In this example as shown, five P-mode edges
exist, based on the light and dark regions in the
measuring area.
Target center pitch: The space between 2P and
4P
Bottom
Center of
optical axis
Top
1E
Edge 0
2E
35.5
±
0.2
Edge position of workpiece
: Space between+1E and 0
(+1E to 0E)
In the edge 0 method, the E and
P modes will not make any difference.
Tips