One-point calibration settings – KEYENCE LS-7600 User Manual
Page 89

5-27
Chapter 5 Function Settings
5
6
Press the [ENT] key.
The input value is entered and the screen returns to the AREA selection status of
calibration settings.
Note:
If the calibration conditions are not satisfied, ERR1 will be displayed. Press
the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the
calibration settings again.
Tips
When repeating the two-point calibration, the previous settings will appear on
the screen, which will be different from those in the above setting example.
One-point Calibration Settings
Make one-point calibration settings as explained below.
Note:
In the case of one-point calibration, the number of averaging measurements is
fixed at 4,096 times.
(1) Press the [Up or Down] key and
select + or – as desired.
(2) Press the [Left or Right] key and
move the cursor between digits to
input the value.
Press the [ENT] key when the cursor is at 1-TARGET.
1
Place the master workpiece in the beam.
2
While the cursor is at T1-B, press the [ENT] key. Monitor the measured value
and input the T1-B value.
The display “*********” changes to “+00.00000.”
The cursor moves to the “+” position and to the value can be input.
AREA1
1-TARGET
T1-A
0 .00000
T1-B
+
00 .00000
T2-B
✽✽✽✽✽✽✽✽✽
SET
5
0 0 3 0 0
AREA1
1-TARGET
T1-A
0 .00000
T1-B
+0
5
.00000
T2-B
✽✽✽✽✽✽✽✽✽
SET
5
0 0 3 0 0
(3) Press the [Up or Down] key to
input the value.
(4) Repeat steps (2) and (3) until all
the digits are input.
(5) Press the [ENT] key.
When the input value is entered, the
measured value of the master
workpiece is set for the T1-A. Then
the cursor moves to T1-B.
4
Press the [Down] key and move teh cursor to SET.