One-point calibration settings – KEYENCE LS-7600 User Manual

Page 89

background image

5-27

Chapter 5 Function Settings

5

6

Press the [ENT] key.

The input value is entered and the screen returns to the AREA selection status of
calibration settings.

Note:

If the calibration conditions are not satisfied, ERR1 will be displayed. Press
the [ENT] key or [ESCAPE] key to reset the error status. Then repeat the
calibration settings again.

Tips

When repeating the two-point calibration, the previous settings will appear on

the screen, which will be different from those in the above setting example.

One-point Calibration Settings

Make one-point calibration settings as explained below.

Note:

In the case of one-point calibration, the number of averaging measurements is
fixed at 4,096 times.

(1) Press the [Up or Down] key and

select + or – as desired.

(2) Press the [Left or Right] key and

move the cursor between digits to
input the value.

Press the [ENT] key when the cursor is at 1-TARGET.

1

Place the master workpiece in the beam.

2

While the cursor is at T1-B, press the [ENT] key. Monitor the measured value
and input the T1-B value.

The display “*********” changes to “+00.00000.”
The cursor moves to the “+” position and to the value can be input.

AREA1

1-TARGET

T1-A

0 .00000

T1-B

+

00 .00000

T2-B

✽✽✽✽✽✽✽✽✽

SET

5

0 0 3 0 0

AREA1

1-TARGET

T1-A

0 .00000

T1-B

+0

5

.00000

T2-B

✽✽✽✽✽✽✽✽✽

SET

5

0 0 3 0 0

(3) Press the [Up or Down] key to

input the value.

(4) Repeat steps (2) and (3) until all

the digits are input.

(5) Press the [ENT] key.

When the input value is entered, the
measured value of the master
workpiece is set for the T1-A. Then
the cursor moves to T1-B.

4

Press the [Down] key and move teh cursor to SET.