Minimum input period, Group processing timing chart, Statistical processing timing chart – KEYENCE LS-7600 User Manual
Page 162: Chapter 7 i/o terminals

7-14
Chapter 7 I/O Terminals
7
■
Minimum Input Period
Refer to the relevant timing charts for the delay time (response delay time) in response
to input.
Name
Specifications
TIMING/ZERO/RESET
P1~P4
ON : 2.1ms
OFF : 2.1ms
ON : 6.0ms
OFF : 6.0ms
■
Group Processing Timing Chart
ON
OFF
OUT2 BCD output
ON
OFF
OUT2 STROBE output
ON
OFF
OUT2 RESET input
OUT1 Measured value
ON
OFF
OUT1 STROBE output
OUT2 WAITING output
ON
OFF
ON
OFF
OUT2 Comparator output
1
n-1
n
1
1
2
2
2
Discrimination result of group A
Group A
Group B
Discrimination result of group A
Discrimination result of group A
Max2.5ms
Min2.1ms
Min2.1ms
STROBE time
0
±
0.1ms
• The result of group discrimination (OUT2) will be determined and output when the set number (n) of OUT1
measured values is determined.
• STROBE time settings can be changed.
Refer to Strobe Output Time on page 5-47.
• OUT2 TIMING input and ZERO input are ignored.
■
Statistical Processing Timing Chart
Statistical results
Measured value
ON
OFF
Strobe output
STATS RUN output
ON
OFF
ON
OFF
STATS input
n-1
n
1
2
Statistical result of 1 to n
Max
Statistical data sampling period
2.5ms
Max
15ms
Max
2.5ms
• If the LS-7600 Series is under external synchronous control, sampling will start with the first determined
data after the rising of STATS input is recognized. The sampling operation of the LS-7600 Series will
continue up to the determined data at the moment the falling of STATS input is recognized, and the whole
data will be statistically processed. On completion of statistical processing, STATS RUN output will turn
OFF.
• If the LS-7600 Series is under internal synchronous control, sampling will start with the first determined
data after the rising of STATS input is recognized. The number of data items that have been preset will be
statistically processed. The statistical processing of the data starts after the number of data items is
determined. On completion of statistical processing, STATS RUN output will turn OFF. If the STATS input is
turned OFF while the data is sampled, the LS-7600 Series will cancel the sampling operation. Statistical
processing is unavailable if STATS input is turned OFF before the LS-7600 Series starts sampling data.