KEYENCE LS-7600 User Manual
Page 104

5-42
Chapter 5 Function Settings
5
Setting Method
1
Press [ENT] key (Down key) and move the cursor to MEA-MODE.
2
Press the [ENT] key.
The cursor moves to the setting item.
OUT1
CALC
+1
A1
+0
A2
AVE
512
MEA-MODE SELF-T 0
1
00
OFFSET
0.00000
3
Press the [Up or Down] key and
select the MEA-MODE from the
following list.
OUT1
CALC
+1
A1
+0
A2
AVE
512
MEA-MODE
NORMAL
OFFSET
0.00000
The following table lists available setting items with corresponding displays on the LCD
monitor.
Normal
Peak hold
Bottom hold
Peak to peak hold
Average hold
Auto peak hold
NORMAL
P-HOLD
B-HOLD
D-HOLD
AV-HOLD
A-P-HOLD
Setting
LCD Monitor
Auto bottom hold
Auto peak to peak hold
Sample hold 1
Sample hold 2
Self timing hold
Setting
LCD Monitor
A-B-HOLD
A-D-HOLD
S-HOLD1
S-HOLD2
SELF-T
■
Self-timing Hold Setting
Select SELF-T and input the self-timing period.
The possible setting range is from 1 to 9999 ms.
(1) Press the [ENT] key.
The cursor moves to the 0 and numeric input will be possible.
(2) Press the [Left or Right] key and move the cursor between digits to input
the value.
(3) Press the [Up or Down] key and
select the value.
(4) Repeat steps (2) and (3) until all
the digits are input.
4
Press the [ENT] key.
All the input items are entered and the cursor returns to MEA-MODE.