Limit range – KEYENCE LS-7600 User Manual
Page 112

5-50
Chapter 5 Function Settings
5
■
Setting OUT1 Output
(1) Press the [ENT] key.
■
Setting OUT2 Output
(1) Press the [Down] key and move
the cursor frame to OUT2.
(2) Press the [ENT] key.
Limit Range
The following ranges are used for the limits.
■
Reference Value
HI output range: (Reference value + Upper limit) < Measured value
GO output range: (Reference value + Lower limit) <
= Measured value
<
= (Reference value
+ Upper limit)
LO output range: Measured value < (Reference value + Lower limit)
Possible Setting Conditions
A possible set value must be within a range where the following two items can be set.
Reference value + Upper limit value
Reference value + Lower limit value
In addition, the following condition must be satisfied.
Reference value + Upper limit value > Reference value + Lower limit value
Tips
Refer to Units, Default Minimum Values, and Setting Ranges on page 5-60 for
possible setting ranges in detail.
■
Threshold
HH output range: HH set value < Measured value
HI output range: HI set value < Measured value <
= HH set value
GO output range: LO set value <
= Measured value
<
= HI set value
LO output range: LL set value <
= Measured value < LO set value
LL output range: Measured value < LL set value
Possible Setting Conditions
HH set value > HI set value > LO set value > LL set value
Tips
Refer to Units, Default Minimum Values, and Setting Ranges on page 5-60 for
possible setting ranges in detail.
OUT1
UPPER
1.00000
STANDARD
10.00000
LOWER
—1.00000
OUT2
UPPER
1.00000
STANDARD
10.00000
LOWER
—1.00000
CAM1
PROG
P — 00
LIMIT
AREA
OUTPUT
CALIB
OPTION
A1 2
A2 2
01/02/18
17:47:30