Chapter 6 Environment Settings
6
3
Press the [Up or Down] key and select the data mode.
The following table lists available setting items with corresponding displays on the
LCD monitor.
Only the item OFF is selectable in data transmission mode.
Description
Setting item
NORMAL
The items S1, S2, or S3 is selectable in data transmission mode.
PRINTER1/PRINTER2
The following environment settings are fixed.
Baud rate: 9600
Data length: 8
Stop bit length: 1
Parity: Even
PLC
4
Press the [ENT] key.
The input item is entered and the cursor returns to D-MODE.
■
Data Transmission Mode
1
Press the [Up or Down] key and select D-SEND.
2
Press the [ENT] key.
The cursor moves to the setting item.
3
Press the [Up or Down] key and select the data transmission mode.
4
Press the [ENT] key.
The input item is entered and the cursor returns to D-SEND.
Head
Take the following steps to set the LED inspection function and mutual interference preven-
tion function.
Setting Item
• LED inspection function:
The intensity of the LED light source can be
inspected.
• Mutual interference prevention function: Prevents mutual interference that may result
when two measuring heads are located at
right angles.
■
LED Inspection Function
Use this function to check for change of the LED light source or the fluctuation of light
intensity caused by dirt on the cover glass. It is recommended to make a regular check
on the change of light intensity after the installation of the LS-7600 Series.