Two-point calibration settings – KEYENCE LS-7600 User Manual
Page 88

5-26
Chapter 5 Function Settings
5
Two-point Calibration Settings
Make two-point calibration settings as explained below.
Setting Range
Two-point calibration requires two types of master workpieces that are different to each
other in size. If the master workpieces are target 1 (T1) and target 2 (T2) respectively,
make settings so that the following conditions will be satisfied.
T1-A < T2-A
Furthermore, set the calibration span between 0.5 and 2.0.
Note:
In the case of two-point calibration, the number of averaging measurements is
fixed at 4,096 times.
Press the [ENT] key when the cursor is at 2-TARGET.
1
Place the small master workpiece (T1-A) in the beam.
2
While the cursor is at T1-B, press the [ENT] key. Monitor the measured value
and input the T1-B value.
The display “*********” changes to “+00.00000.”
The cursor moves to the “+” position and to the value can be input.
(1) Press the [Up or Down] key and
select + or – as desired.
(2) Press the [Left or Right] key and
move the cursor between digits to
input the value.
AREA1
2-TRAGET
T1-A
0 .00000
T1-B
+
00 .00000
T2-A
✽✽✽✽✽✽✽✽✽
T2-B
✽✽✽✽✽✽✽✽✽ SET
5
0 0 3 0 0
AREA1
2-TRAGET
T1-A
0 .00000
T1-B
+0
5
.00000
T2-A
✽✽✽✽✽✽✽✽✽
T2-B
✽✽✽✽✽✽✽✽✽ SET
5
0 0 3 0 0
(3) Press the [Up or Down] key to
input the value.
(4) Repeat steps (2) and (3) until all
the digits are input.
(5) Press the [ENT] key.
When the input value is entered, the
measured value of the master
workpiece is set for the T1-A. Then
the cursor moves to T1-B.
3
Place the large master workpiece (T2-A) in the beam.
4
Press the [Down] key and select T2-B. Then input the value in the same man-
ner in step 2.
5
Press the [Down] key and move the cursor to SET.