Tolerance mode (limits), Timing mode – KEYENCE LS-7600 User Manual
Page 118

5-56
Chapter 5 Function Settings
5
Tolerance Mode (LIMITS)
Take the following steps to select the appropriate tolerance mode.
Setting Range
The following modes are selectable.
• Reference value mode: Use this mode if there is a workpiece size change with the
deviations (tolerances) unchanged.
• Threshold mode:
Specify the upper and lower limits in absolute values.
1
Press the [Up or Down] key and move the cursor to LIMITS.
2
Press the [ENT] key.
The cursor moves to the setting item.
3
Press the [Up or Down] key and
select the tolerance mode.
The following table lists available setting
items with corresponding displays on the
LCD monitor.
Reference value mode
DEV
Setting
LCD monitor
Threshold mode
TOL
4
Press the [ENT] key.
The input item is entered and the cursor
returns to LIMITS.
LIMITS
DEV
T-MODE
ASYNC
D-UNIT
mm
D-RES
0.00005
C-OUT
SUB
GRAPH-T
OUT1
1m
OUT2
1m
STATS
OUT1
OFF
OUT2
OFF
GROUP
OFF
APP-MODE OFF
Timing Mode
Take the following steps to set the synchronous (timing) input terminals corresponding to
OUT1 and OUT2 respectively.
Setting Range
Select the input terminals from the following ones.
• Asynchronous mode (ASYNC): The LS-7600 Series is in asynchronous control of
OUT1 and OUT2. Select the input terminals independ-
ently for OUT1 and OUT2 respectively. Terminal
number 13 corresponds to OUT1 and connector I/O
number 38 corresponds to OUT2.
• Synchronous mode (SYNC):
The LS-7600 Series is in synchronous control of
OUT1 and OUT2. Terminal number 13 corresponds to
them. Connector I/O number 38 is disabled.
Refer to Section 7 I/O Terminals on page 7-1 for details.