KEYENCE LS-7600 User Manual
Page 115

5-53
Chapter 5 Function Settings
5
■
Setting Upper Limit Hold Value
Take the following steps to set the upper limit hold value so that the LS-7600 Series will
automatically detect and eliminate measured values exceeding the upper limit hold
value. This function is enabled with the set value turned ON and disabled with the set
value turned OFF.
●
Turning OFF
1
Press the [Up or Down] key and move the cursor to HOLD-H.
2
Press the [ENT] key.
The cursor moves to the setting item.
3
Press the [Up or Down] key and
select OFF.
4
Press the [ENT] key.
The input item is entered and the cursor moves to HOLD-H.
OUT1
HOLD—H
OFF
HOLD—L
OFF
●
Turning ON
1
Press the [Up or Down] key and move the cursor to HOLD-H.
2
Press the [ENT] key.
The cursor moves to the setting item.
OUT1
HOLD—H
ON
+
90.00000
HOLD—L
OFF
3
Press the [Up or Down] key and
select ON.
The set value is displayed below the
display “ON.”
4
Press the [ENT] key.
The set value is displayed below the
display “ON.”
5
Press the [ENT] key.
The cursor moves to +, where numeric
input is possible.
OUT1
HOLD—H
ON
+
90.00000
HOLD—L
OFF
6
Press the [Up or Down] key and
select + or –.
7
Press the [Left or Right] key and
move the cursor to the digit where the
value is input.